Temperature
Working-temperature range |
5 °C above ambient temperature up to +50 °C
Standard sterilisation programme: 60 minutes at 180°C (without removing the sensors) |
Setting temperature range |
+18 to +50 °C |
Setting accuracy temperature |
0.1 °C |
Temperature sensor |
2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error |
Temperature variation in chamber |
at + 37 °C +/- 0.3 K |
Temperature fluctuations with time |
at 37 °C +/- 0.1 K |
Humidity
Humidity control (standard) |
Humidity limitation thanks to a Peltier element; when water dish is full and inserted, the Peltier element limits the value of relative humidity in the interior to 93 % rh +/- 2.5 % |
Setting accuracy humidity |
0.5 % rh |
Setting range active humidity control (with option K7) |
40 to 97 % rh and rh-Off |
Control of standard components
CO2 control |
Digital electronic CO2 control with dual beam NDIR system, with auto-diagnostic system and acoustic fault indication, barometric pressure compensation |
Adjustment range CO2 |
0 to 20 % CO2 |
Setting accuracy CO2 |
0,1% |
Variation in time CO2 |
+/- 0.2 % CO2 |
Adjustment range O2 |
1 to 20 % O2 |
Setting accuracy O2 |
0.1 % O2 |
Control technology
ControlCOCKPIT |
TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays. |
Language setting |
German, English, Spanish, French, Polish, Czech, Hungarian |
Function SetpointWAIT |
the process time does not start until the set temperature is reached |
Adjustable parameters |
temperature (Celsius or Fahrenheit), CO2, programme time, time zones, summertime/wintertime |
Communication
Interface |
Ethernet LAN, USB |
Documentation |
programme stored in case of power failure |
Programming |
AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port |
Safety
AutoSAFETY |
additionally integrated over- and undertemperature protection “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating function is switched off in case of overtemperature, cooling function in case of undertemperature |
Autodiagnostic system |
integral fault diagnostics for temperature and CO2 |
Alarm |
visual and acoustic |
Heating concept
6 sides |
large-area multi-function heating system on four sides with additional door and back heating to avoid condensation |
This product is already in your quote request list.